TERUMO CARDIOVASCULAR SYSTEMS CORP. LARGE (6) ROLLER PUMP FOR TERUMO ADVANCED PERFUSION SYSTEM 1; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ 6 INCH ROLLER PUMP FOR ADVANC
|
Back to Search Results |
|
Model Number 816571 |
Device Problems
Device Stops Intermittently (1599); Device Displays Incorrect Message (2591)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/22/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Evaluation is in progress, but not yet concluded.As per field service representative (fsr), the facility's biomed is terumo trained and does the in-house repairs for this hospital.The biomed has already inspected the unit.This complaint is related to (b)(4) medwatch #1828100-2017-00015 and 1828100-2016-00817.
|
|
Event Description
|
It was reported that during priming of the device for a cardiopulmonary bypass (cpb) procedure, the roller pump stopped working and the service light illuminated.A service pump message was displayed on the arterial pumps at the time of the pump stop.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.During laboratory analysis, the product surveillance technician (pst) could not duplicate the reported complaint.He operated the pump for approximately 180 hours with no pump jam and no "service pump" message observed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|