Catalog Number 227435500
Device Problem Break (1069)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Date 01/06/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The product associated with this reported event was not returned for examination.Requests for additional investigational inputs were made in accordance with (b)(4) appendix a.No additional information was obtained.A complaint database search has identified a trend for this failure within the 2274 quickset drill family.Previous investigations have determined that use error is the likely root cause.As a result of trending health hazard evaluation, (b)(4) was conducted.The investigation did not establish the need for corrective action.No evidence was found of product or design error as a contributing factor.Complaints will be monitored under (b)(4) post market surveillance.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Event Description
While drilling to put a screw into the cup, the end of the drill bit broke and was left in the patient's bone.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUICKSET 1PC FLEX DRILL BIT 35
Type of Device
HIP INSTRUMENT/TRIAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC.
700 orthopaedic drive
warsaw IN 46582
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC.
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
chad gibson
700 orthopaedic drive
warsaw, IN 46581
5743725905
MDR Report Key6258610
MDR Text Key65029293
Report Number1818910-2017-11084
Device Sequence Number1
Product Code HWT
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Physician
Type of Report Initial
Report Date 01/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number227435500
Device Lot NumberPG243794
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/06/2017
Initial Date FDA Received01/17/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/08/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-