Catalog Number 10-401
Device Problem Metal Shedding Debris (1804)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/14/2016
Event Type  malfunction  
Manufacturer Narrative
Event problem and evaluation-the disposable product was returned.Gross visual analysis performed and revealed a chip on the blade assembly edge, but no fragmentation of blade edge was noted.The cutting window rim was clean indicating no bade exit occurred during procedure.No blade assembly fragments were found in return packaging.The cause of blade chip is unknown.All myosure disposables are tested for proper mechanical function and blade damage prior to qa release.This observation will be monitored and trended.Device history record (dhr) review was conducted for the reported identification number.The lot was released meeting all qa specifications.Currently unable to establish a relationship or impact to the reported observation.(b)(4).
 
Event Description
It was reported during a myosure procedure for uterine tissue removal on (b)(6) 2016, the physician noticed metal shavings in the patient and removed it with graspers.No patient injury and the patient was discharged home.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MYOSURE HYSTEROSCOPIC TISSUE REMOVAL SYSTEM
Type of Device
UTERINE TISSUE REMOVAL SYSTEM
Manufacturer (Section D)
HOLOGIC, MARLBORUGH MA
250 campus drive
marlborough MA 01752
Manufacturer Contact
sidra piracha
250 campus drive
marlborough, MA 01752
5082638884
MDR Report Key6260632
MDR Text Key65576505
Report Number1222780-2017-00005
Device Sequence Number1
Product Code HIH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K100559
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 12/19/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date05/30/2019
Device Catalogue Number10-401
Device Lot Number16E30R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/19/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/19/2016
Initial Date FDA Received01/18/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/08/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MYOSURE CONTROL UNIT-SERIAL # UNK; MYOSURE HYSTEROSCOPE-SERIAL # UNK
-
-