The manufacturing batch lot was not provided therefore we were unable to review the device history records to confirm that the device met all material, assembly and inspection specifications prior to shipment.The device was received for analysis on 01/12/2017.As received, the specimen consisted of one each clinically used, damaged unidentified safari wire; returned coiled, loose and double-bagged within "zip-lock" style poly biohazard pouches.The specimen presented a cut through the coil and core wires 238.35cm from the proximal end; all material distal of the cut was missing.The specimen also presented several bends/kinks of varying severity and frequency scattered over the length of the device beginning 244.7cm from the proximal aspect of the coil and core wire cut, and scraped/frayed ptfe coating scattered over the length of the device.There was no mention of the guidewire being cut in the complaint indicating it was cut in order to remove it from the lotus system.At this time, it is not possible to assign a definitive root cause for the event as reported.Based on the evidence presented by the sample and the information provided by the supporting documentation, clinical and/or procedural factors appear to have impacted on the event as reported.If additional information, or lot traceability is received a follow-up medwatch report will be filed.
|