Catalog Number 26-1221
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Information (3190)
Event Date 12/21/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Upon completion of the investigation, it was noted that the evaluation of the returned perforator revealed dark spots on the cutting area (metal surfaces, 3 locations).A visual inspection for particulate was performed.The particles on this perforator are smaller (or same size) as the test acceptable size requirements.The returned perforator was found to meet the packaging specifications.A review of the dhr revealed that this perforator met all testing requirements during manufacturing processes.This complaint is considered to be closed based on the particulates meeting specification.Based on the results of this investigation, no further action is required.Trends will be monitored for this and similar complaints.At the present time, this complaint is closed.
 
Event Description
Foreign substances were found inside the brand new package before use.There was no adverse consequence to the patient and no further information was provided by the hospital.
 
Manufacturer Narrative
Upon completion of the investigation it was noted that the evaluation of the returned perforator revealed staining on the cutting surfaces of the perforator.The exact root cause of the staining could not be verified; however, this staining is consistent in appearance with previous staining evaluated by the supplier.Previous evaluation of staining on returned perforators (found before use) is most likely associated with water residue attributed to the rinse process, this however could not be determined.The perforator for this complaint was manufactured prior to the changes implemented in july 2016.Dhr review was performed and revealed that this perforator meets all testing requirements during the manufacturing process.Further evaluation of the perforator for this complaint (perforator lot hj011s) is ongoing to address the affiliate's question regarding biocompatibility.If the evaluation reveals any additional information, this complaint will be reopened and a follow up will be filed.Based on the results of this investigation no further action is required.Trends will be monitored for this and similar complaints.At the present time this complaint is closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN DISPOSABLE PERFORATOR
Type of Device
DRILLS, BURRS, TREPHINES & ACC.
Manufacturer (Section D)
CODMAN & SHURTLEFF, INC.
new bedford industrial park
new bedford MA 02745
Manufacturer Contact
james kenney
325 paramount drive
raynham, MA 02767
5088282726
MDR Report Key6260784
MDR Text Key65098772
Report Number1226348-2017-10055
Device Sequence Number1
Product Code HBF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K791101
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number26-1221
Device Lot NumberHJ011S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/16/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/21/2016
Initial Date FDA Received01/18/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received01/26/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-