(b)(4).The complaint airvo humidifier was not returned to fisher & paykel healthcare (fph) for evaluation as it continues to be used by the hospital.The hospital confirmed that the airvo correctly alarmed to inform staff that the set oxygen levels had dropped.Conclusion: the 900pt402 oxygen inlet kit is supplied with the airvo 2 humidifier and allows the user to connect supplementary oxygen via a barb connector, which is suitable for use with most common sizes of oxygen tubing.The hospital did not confirm what had caused the oxygen tubing to become disconnected.No failure of the subject airvo 2 was alleged and we can conclude that it continues to deliver therapy as intended.The airvo user manual states that the "airvo is for the treatment of spontaneously breathing patients who would benefit from receiving high-flow, warmed and humidified respiratory gases." and that " the unit is not intended for life support." the airvo 2 user manual instructs the user to "use continuous oxygen monitoring on patients who would desaturate significantly in the event of disruption to their oxygen supply." the airvo 2 product technical manual informs the user how to set the low oxygen alarm and states: "when set to 25% the unit will alarm if the measured oxygen fraction is below this value.This allows detection of oxygen being disconnected.".
|