• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANIMAS CORPORATION ANIMAS VIBE; INSULIN INFUSION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANIMAS CORPORATION ANIMAS VIBE; INSULIN INFUSION PUMP Back to Search Results
Device Problems No Display/Image (1183); Structural Problem (2506)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The pump has not been returned to animas at the time of this report.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
 
Event Description
On (b)(6) 2016, a reporter contacted animas alleging that the display screen was cracked.This complaint is being reported because the issue may impact the user's ability to read some or all of the information on the screen which may result in over or under delivery.There was no serious injury associated with this complaint.
 
Manufacturer Narrative
The device has been returned and evaluated by product analysis on 02/08/2017 with the following findings: animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.During a visual inspection of the pump, a superficial crack was found near the perimeter of the display lens.The complaint was duplicated.Unrelated to the complaint, the battery compartment was observed to be cracked.Additionally, the bolus button cover was torn; however, the button responded properly to user presses.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Type of Device
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
Manufacturer (Section G)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
Manufacturer Contact
karin sargrad
200 lawrence dr
west chester, PA 19380-3428
4843561808
MDR Report Key6262706
MDR Text Key65508122
Report Number2531779-2017-01541
Device Sequence Number1
Product Code MDS
Combination Product (y/n)N
PMA/PMN Number
P130007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Type of Report Initial,Followup
Report Date 12/23/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/03/2017
Was the Report Sent to FDA? Yes
Initial Date Manufacturer Received 12/23/2016
Initial Date FDA Received01/18/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received03/02/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-