• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT OMNILINK ELITE PERIPHERAL STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT OMNILINK ELITE PERIPHERAL STENT SYSTEM Back to Search Results
Catalog Number 1012625-29
Device Problems Improper or Incorrect Procedure or Method (2017); Physical Resistance (2578); Device Damaged by Another Device (2915)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/13/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.(b)(4).The device was not returned for evaluation.A review of the lot history record and complaint history of the reported lot could not be conducted, because the lot number was not provided.It should be noted that the omni elite instructions for use states: when treating multiple lesions, the distal lesion should be initially stented, followed by stenting of the proximal lesion.Stenting in this order obviates the need to cross the proximal stent in placement of the distal stent and reduces the chances for dislodging the stent.The investigation determined the reported difficulties to be related to user error as a deviation to the ifu contributed to the reported difficulties.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.The other omnilink elite referenced in describe event or problem and concomitant medical products is being filed under a separate medwatch report number.
 
Event Description
It was reported that the procedure was to treat a moderately calcified right external iliac artery.It was a contralateral approach.A 9.0 x 29 mm omnilink elite stent was deployed in the lesion.A second 9.0 x 29 mm omnilink elite stent delivery system was being advanced through the previously placed stent when there was resistance with the tip and the deployed stent, and the deployed stent compressed approximately 10 mm at the proximal end.The second stent was deployed distally to the first stent.A third stent was deployed proximal to the first stent to cover the area where the stent compressed.There was no clinically significant delay in procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNILINK ELITE PERIPHERAL STENT SYSTEM
Type of Device
PERIPHERAL STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, TEMECULA, CA USA REG# 2024168
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key6265314
MDR Text Key65578487
Report Number2024168-2017-00574
Device Sequence Number1
Product Code NIO
UDI-Device Identifier08717648179334
UDI-Public(01)08717648179334
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P110043
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 01/19/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number1012625-29
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/04/2017
Initial Date FDA Received01/19/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
STENT: OMNILINK ELITE
Patient Age75 YR
-
-