• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN AVITUM AG - MELSUNGEN DIALOG A + HE/BIC; HEMODIALYSIS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN AVITUM AG - MELSUNGEN DIALOG A + HE/BIC; HEMODIALYSIS SYSTEM Back to Search Results
Catalog Number 710200L
Device Problem Use of Device Problem (1670)
Patient Problem Cardiac Arrest (1762)
Event Date 01/20/2017
Event Type  Death  
Manufacturer Narrative
(b)(4).The machine trend file and additional information has been requested from the facility and the investigation is on-going at this time.A follow up report will be provided when the evaluation results are available.
 
Event Description
As reported by the user facility: customer reported that patient had died while receiving dialysis treatment.A bbraun technician spoke to the hospital biomed and learned that the incident occurred around 4pm.The biomed reported that he did not have details about the incident.A follow up call was made with the facility risk management.Risk management representative indicated that the patient suffered blood loss due to the venous needle coming dislodged from the access site.She did not know how much blood was lost.She indicated that cpr was administered to the patient, but the patient died on site.The dialysis machine was sequestered, waiting for bbraun to conduct a machine inspection.
 
Manufacturer Narrative
(b)(4).Customer facility reported an incident related to the death of a patient, while receiving dialysis therapy on the dialog+ dialysis machine (sn (b)(4)).According to the customer's report, the patient started her hemodialysis therapy on (b)(6) 2017 at 10:39 am with a pre-dialysis weight of (b)(6).The dialog+ hemodialysis machine was next to the patient's bed.There was no family or visitor present.The patient was alert, oriented and not restless, but it was noticed that she moved the blanket onto her chest frequently during this treatment.The facility reported that the nurses check the patients every 15 minutes during therapy.About two hours into therapy, the nurse was last in the patient's room.The patient was sitting upright in bed talking on the phone.At this time the catheter connection was intact and the patient was normotensive.About 2-5 minutes after the nurse had left the patient's room, she heard a noise.The patient had called out or screamed.When the nurse ran into the room, she noticed the blood on the floor and the patient remained in the bed.It was noticed at that time that the venous access was detached from the blood line set.The luer-lock connectors of the blood line and the catheter were not cracked nor did they appear broken.A rapid response team was immediately at the patient's side, which quickly escalated to code blue team assisting to save the patient's life.Fluids were administered to the patient before the pulse stopped.The patient lost about 2 l of blood within approximately 2.5 min.Cpr started 13:14 pm and was terminated at 13:45 pm.The facility reported that no alarms were triggered by the dialog+ machine when the pressure dropped due to the disconnection of the venous blood line from the catheter.A b.Braun (b)(4) technician was sent to the facility to inspect the machine.According to his report, the technical inspection of the dialog+ dialysis machine did not show any product deviations or malfunctions.The dialog+ hemodialysis machine operated as intended.All information associated with this event was forwarded to the equipment manufacturer b.Braun (b)(4).Also submitted to (b)(4) was the machine trend file associated with this specific therapy.According to the manufacturer, the analysis of the data record of the dialog+ machine showed that 2 hours 20 minutes into therapy the alarm "venous pressure - lower limit - check access" was triggered caused by a sudden drop of the venous pressure from about +158 mmhg to about +40 mmhg.This pressure drop was most probably caused by the disconnection of the venous blood line from the patient's permanent cvc.As a consequence of the alarm, the dialysis machine stopped the blood pump, closed the safety air clamp (sakv) and switched into bypass mode (patient-safe condition).The time between the pressure drop and the alarm and the stop of the blood pump was 5 seconds.During this time the blood loss was calculated to be 33 ml.Twenty-two seconds later the alarm was muted and acknowledged actively by the patient/user by pressing the aq key twice.The analysis of the data trend shows that the machine was started again actively by the patient/user.The evaluation of the trend data record of the dialog+ dialysis machine correlates chronologically with the customer's description of the event.It shows the following sequence of events: the therapy started at 10:39; a drop of the venous pressure at 12:59 lead 5 seconds later to the alarm "venous pressure - lower limit - check access"; as a consequence of the alarm, the blood pump stopped, the safety air clamp (sakv) closed and the device switched into bypass; another 22 seconds later the alarm was muted and acknowledged actively by the patient/user by pressing the aq key twice; cpr was started 14 minutes later.In conclusion, the investigation showed that there was no product deviation.The dialog+ dialysis machine alarmed the drop in the venous pressure as intended, the blood pump stopped, the sakv closed and the machine switched into bypass.All information concerning this incident has been included in our trend analysis of the product line.A historical review of the customer complaint database, revealed no adverse trends regarding this issue.If additional pertinent information becomes available a follow up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIALOG A + HE/BIC
Type of Device
HEMODIALYSIS SYSTEM
Manufacturer (Section D)
B. BRAUN AVITUM AG - MELSUNGEN
schwarzenberger weg 73-79
melsungen, D-342 12
GM  D-34212
Manufacturer (Section G)
B. BRAUN AVITUM AG - MELSUNGEN
schwarzenberger weg 73-79
melsungen, D-342 12
GM   D-34212
Manufacturer Contact
iris ratke
buschberg 1
melsungen, 34212
GM   34212
661713718
MDR Report Key6275676
MDR Text Key65688789
Report Number3002879653-2017-00002
Device Sequence Number1
Product Code FKJ
UDI-Device Identifier040469636
UDI-Public(01)040469636
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K083460
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Risk Manager
Type of Report Initial,Followup
Report Date 03/27/2017,01/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number710200L
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Report Sent to FDA03/27/2017
Distributor Facility Aware Date01/20/2017
Event Location Outpatient Treatment Facility
Date Report to Manufacturer03/27/2017
Initial Date Manufacturer Received 01/20/2017
Initial Date FDA Received01/25/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received03/27/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Death; Other; Required Intervention;
Patient Age50 YR
Patient Weight63
-
-