MAUDE Adverse Event Report: STRYKER INSTRUMENTS-KALAMAZOO WIRE COLLET (.7-1.8MM); ARTHROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-KALAMAZOO WIRE COLLET (.7-1.8MM); ARTHROSCOPE Back to Search Results
Catalog Number 4100062000
Device Problems Overheating of Device (1437); Metal Shedding Debris (1804)
Patient Problem No Patient Involvement (2645)
Event Date 09/01/2016
Event Type  malfunction  
Manufacturer Narrative
This mdr report is part of the malfunction summary reporting program, (b)(4).One device is available for evaluation but has not yet been evaluated.There were no remedial actions taken.This device is not labeled for single-use.
 
Event Description
This report summarizes 1 malfunction event, in which the device was reportedly overheating and had metal shavings coming from it.There was no patient involvement; no patient impact.
 
Manufacturer Narrative
(b)(4).1 device was received for evaluation.The event was confirmed for metal shedding debris.Though the overheating was not duplicated; the device was found to be outside specifications for the reported event.The device was found to have debris, corrosion, and shattered bearings.This device is not repairable and was not returned to the user facility.There were no remedial actions taken.This device is not labeled for single-use.
 
Event Description
This report summarizes 1 malfunction event, in which the device was reportedly overheating and had metal shavings coming from it.There was no patient involvement; no patient impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WIRE COLLET (.7-1.8MM)
Type of Device
ARTHROSCOPE
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer (Section G)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer Contact
casey metzger
4100 east milham avenue
kalamazoo, MI 49001
2693237700
MDR Report Key6280088
MDR Text Key66038118
Report Number0001811755-2017-00271
Device Sequence Number1
Product Code HRX
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)Y
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/31/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number4100062000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/01/2016
Initial Date FDA Received01/26/2017
Supplement Dates Manufacturer Received09/01/2016
Supplement Dates FDA Received07/31/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-