Catalog Number 4100062000 |
Device Problems
Overheating of Device (1437); Metal Shedding Debris (1804)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 09/01/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
This mdr report is part of the malfunction summary reporting program, (b)(4).One device is available for evaluation but has not yet been evaluated.There were no remedial actions taken.This device is not labeled for single-use.
|
|
Event Description
|
This report summarizes 1 malfunction event, in which the device was reportedly overheating and had metal shavings coming from it.There was no patient involvement; no patient impact.
|
|
Manufacturer Narrative
|
(b)(4).1 device was received for evaluation.The event was confirmed for metal shedding debris.Though the overheating was not duplicated; the device was found to be outside specifications for the reported event.The device was found to have debris, corrosion, and shattered bearings.This device is not repairable and was not returned to the user facility.There were no remedial actions taken.This device is not labeled for single-use.
|
|
Event Description
|
This report summarizes 1 malfunction event, in which the device was reportedly overheating and had metal shavings coming from it.There was no patient involvement; no patient impact.
|
|