SMITH & NEPHEW, INC. LIGHT SOURCE, 500XL, XENON; ILLUMINATOR, FIBEROPTIC, FOR ENDOSCOPE Back to Search Results
Catalog Number 72200568
Device Problem Inadequate Lighting (2957)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/27/2016
Event Type  malfunction  
Event Description
It was reported that during the shoulder procedure, the light source went out.A competitor's back-up was used to complete the procedure.
 
Manufacturer Narrative
Complaint of shutting down was confirmed.Product overheated due to dead ballast fan.Cause of fan malfunction is a defective mcu pcb.Unit passes functional testing and fans perform as expected with a known good mcu pcb installed.After the evaluation the root cause for the reported issue was defective electronic components.A review of the device history record was performed which confirmed no inconsistencies.There are no indications that would suggest that the device did not meet product specifications upon release into distribution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGHT SOURCE, 500XL, XENON
Type of Device
ILLUMINATOR, FIBEROPTIC, FOR ENDOSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
jim gonzales
7000 w. william cannon
austin, TX 78735
5123585706
MDR Report Key6292957
MDR Text Key66510516
Report Number1643264-2017-00063
Device Sequence Number1
Product Code FFS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K994084
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 01/11/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72200568
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/18/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/11/2017
Initial Date FDA Received02/01/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received02/15/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/23/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-