The event unit was returned for evaluation.Upon inspection, engineering confirmed that the latis pad was detached from the one of the jaws.The exact root cause of the fracture is unknown, potential causes could be the result of prolonged exposure in a lipid-rich environment or excessive force.All graspers undergo 100% visual and functional inspection during the manufacturing and assembly process.Applied medical will continue to monitor its vigilance system for trends and take appropriate actions as necessary.This report is being filed as a result of a re-review of applied medical complaints received between (b)(6) 2014 and (b)(6) 2016. this retrospective review was associated with a quality management system (qms) compliance action plan developed and presented to fda to address an april 10, 2015 warning letter.Applied medical has revised its mdr reporting criteria to be more conservative and has improved complaint handling and mdr reporting processes.The reviews ensured that recent reportable events were appropriately identified and reported to the designated regulatory authority(ies).This report, which represents the initial and final reports combined, is being submitted based on the findings of that retrospective review.In accordance to 21 cfr 803.56, if additional information is obtained which was not known or was not available when this report was submitted, then a supplemental report will be submitted to the fda.
|