• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SORIN GROUP DEUTSCHLAND SORIN S5 SYSTEM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SORIN GROUP DEUTSCHLAND SORIN S5 SYSTEM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 48-40-00
Device Problems Device Displays Incorrect Message (2591); Connection Problem (2900)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/02/2017
Event Type  malfunction  
Manufacturer Narrative
Patient information was not provided.Sorin group (b)(4) manufactures the sorin s5 system.The incident occurred in (b)(6).This medwatch report is being filed on behalf of sorin group (b)(4).A sorin group service representative was dispatched to the facility to investigate.The console was unavailable for testing, so the issue could not be confirmed.A serial read out of the cardioplegia module log was performed and a new cardioplegia module was ordered for replacement.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.The investigation is ongoing.A follow-up report will be sent when the investigation is complete.
 
Event Description
Sorin group (b)(4) received a report that pump 4a of the sorin s5 system lost its link to the cardioplegia module during a procedure and was no longer controlled by the cardioplegia module.The cardioplegia module alarmed and an error was displayed on the s5 console.There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SORIN S5 SYSTEM
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM  80939
Manufacturer (Section G)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM   80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key6296620
MDR Text Key66435577
Report Number9611109-2017-00024
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 02/02/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number48-40-00
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/06/2017
Initial Date FDA Received02/02/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/12/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-