• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - MARLBOROUGH SOLYX SIS SYSTEM; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR STRESS URINARY INCONTINENCE, FEMA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - MARLBOROUGH SOLYX SIS SYSTEM; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR STRESS URINARY INCONTINENCE, FEMA Back to Search Results
Model Number M0068507000
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Internal Organ Perforation (1987)
Event Date 01/11/2017
Event Type  Injury  
Manufacturer Narrative
(b)(4).The complainant indicated that the device is disposed and will not be returned for evaluation; therefore a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
 
Event Description
It was reported to boston scientific corporation that a solyx sis system was used during a mid urethral sling placement procedure performed on (b)(6) 2017.According to the complainant, during implantation on the right side, instead of implanting the device in the obturator internus muscle, the device was implanted into the trigone of the bladder.This perforation of the patient's bladder and was identified through cystoscopy.To remove the mesh, he attached a scissor to the cystoscope, cut the right-side anchor out of the mesh, then pulled it out with the scope through the urethra.The remaining parts of the sling were removed vaginally.The left anchor remained in the left obturator internus, but the mesh was cut close to the anchor such that 95% of the mesh was removed.The procedure was completed with another of the same device.A foley catheter was placed to be left in for a week, and the patient will return for a follow up check up in a week.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOLYX SIS SYSTEM
Type of Device
MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR STRESS URINARY INCONTINENCE, FEMA
Manufacturer (Section D)
BOSTON SCIENTIFIC - MARLBOROUGH
100 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC - MARLBOROUGH
100 boston scientific way
marlborough MA 01752
Manufacturer Contact
nancy cutino
100 boston scientific way
marlborough, MA 01752
5086834000
MDR Report Key6298409
MDR Text Key66429413
Report Number3005099803-2017-00202
Device Sequence Number1
Product Code PAH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K081275
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 01/11/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Expiration Date09/14/2019
Device Model NumberM0068507000
Device Catalogue Number850-700
Device Lot Number19717964
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/11/2017
Initial Date FDA Received02/02/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/27/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age44
-
-