• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CORPORATION IMUFLEX BLOOD BAG SYSTEM; IMUFLEX WB-RP BLOOD BAG SYSTEM WITH LR FILTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CORPORATION IMUFLEX BLOOD BAG SYSTEM; IMUFLEX WB-RP BLOOD BAG SYSTEM WITH LR FILTER Back to Search Results
Catalog Number 1BBWGQ506A2
Device Problems High Test Results (2457); Filtration Problem (2941); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Patient Involvement (2645)
Event Date 01/09/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Investigation is still in process.A follow-up report will be provided.
 
Event Description
The customer reported elevated white blood cell (wbc) content in a filtered whole blood unit.There was not a transfusion recipient or patient involved at the time of the unit processing,therefore no patient information is reasonably known at the time of the event.(b)(6).
 
Manufacturer Narrative
Investigation: a red blood cell (rbc) storage bag was received for evaluation.However,an evaluation could not be performed since the filter was not returned for evaluation and the reported event is related to a filtration issue.The manufacturing records, test records, and inspection records were reviewed for abnormalities and none were found.The records regarding the particulate removal rates of the filter membranes were reviewed.All membranes conformed to established specification.Root cause: a definitive root cause for the observed elevated wbc count remains undetermined at this time.Based on the available information, it cannot be ruled out that the higher-than-expected wbc content in the whole blood product could be donor-related(characteristics of blood).It also cannot be ruled out that a filtering error or other process error(not resting the collected blood prior to filtering; not expressing air properly from the product bag) could have contributed to the higher-than-expected wbc content in the whole blood product.The instructions for use provide a caution to not squeeze or apply pressure to the filter while it is attached to the bag containing the filtered blood in order to avoid leukocyte leakage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMUFLEX BLOOD BAG SYSTEM
Type of Device
IMUFLEX WB-RP BLOOD BAG SYSTEM WITH LR FILTER
Manufacturer (Section D)
TERUMO CORPORATION
fujinomiya
Manufacturer Contact
steve kern
10810 w. collins ave
lakewood, CO 80215
3032392246
MDR Report Key6300828
MDR Text Key66871108
Report Number9681839-2017-00023
Device Sequence Number1
Product Code CAK
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 02/17/2017,02/03/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2019
Device Catalogue Number1BBWGQ506A2
Device Lot Number160822KK
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/17/2017
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date01/11/2017
Device Age6 MO
Event Location Hospital
Date Report to Manufacturer02/17/2017
Initial Date Manufacturer Received 01/11/2017
Initial Date FDA Received02/03/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received02/22/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/31/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-