Model Number 37800 |
Device Problem
High impedance (1291)
|
Patient Problem
Therapeutic Effects, Unexpected (2099)
|
Event Type
Injury
|
Manufacturer Narrative
|
Information references the main component of the system and other applicable components are: product id 435135, serial# (b)(4), implanted: (b)(6) 2010, product type: lead.Product id 435135, serial# (b)(4), implanted: (b)(6) 2010, product type: lead.
|
|
Event Description
|
A healthcare provider (hcp) reported that a patient was not getting good therapy.Impedance tests were done and the results indicated "high z on pairs with #2".The results of 2,3 were greater than 4000 ohms.C2 was =>20,000 ohms and c2=354 ohms, 23=>20,000ohms.It was noted that the out of range electrodes were being used in programming and causing therapy problems.It was reviewed with the hcp that they may consider replacing the lead involving #2 as it appeared to have an open circuit.The hcp stated they were doing a surgical procedure that day to address the high impedance and change out the lead.There was no lead migration, per the hcp.The implantable neurostimulator (ins) looked good an appeared to be working.The hcp did state that the implantable neurostimulator (ins) may have depleted more quickly due to the open circuit, but it was reviewed that the open circuits typically don't deplete batteries rapidly.The ins was indicated for gastric stimulation/gastrointestinal/pelvic floor.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|