Model Number R SERIES |
Device Problem
Failure to Select Signal (1582)
|
Patient Problem
Death (1802)
|
Event Type
Death
|
Manufacturer Narrative
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
|
Event Description
|
Complainant alleged that while attempting to treat a patient (age & gender unknown) the device was unable to obtain an ecg signal.Complainant indicated that the clinician obtained another device to continue treating the patient.Complainant indicated that the patient subsequently expired.
|
|
Manufacturer Narrative
|
The device was returned to zoll medical corporation; the malfunction was duplicated and attributed to a lifted pad on a transformer on the analog board.The analog board was unrepairable and replaced as part of the repair.The device was recertified and returned to the customer.The analysis for reports of this type has not identified an increase in trend.
|
|