• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION R SERIES DEFIBRILLATOR; DEFIBRILLATOR/PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION R SERIES DEFIBRILLATOR; DEFIBRILLATOR/PACEMAKER Back to Search Results
Model Number R SERIES
Device Problem Failure to Select Signal (1582)
Patient Problem Death (1802)
Event Type  Death  
Manufacturer Narrative
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
 
Event Description
Complainant alleged that while attempting to treat a patient (age & gender unknown) the device was unable to obtain an ecg signal.Complainant indicated that the clinician obtained another device to continue treating the patient.Complainant indicated that the patient subsequently expired.
 
Manufacturer Narrative
The device was returned to zoll medical corporation; the malfunction was duplicated and attributed to a lifted pad on a transformer on the analog board.The analog board was unrepairable and replaced as part of the repair.The device was recertified and returned to the customer.The analysis for reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R SERIES DEFIBRILLATOR
Type of Device
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key6303471
MDR Text Key66585858
Report Number1220908-2017-00182
Device Sequence Number1
Product Code DRT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K060559
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/17/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Other
Device Model NumberR SERIES
Device Catalogue NumberR SERIES
Other Device ID Number00847946017392
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/02/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/17/2017
Initial Date FDA Received02/06/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/14/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Death;
-
-