Model Number N/A
Device Problems Material Erosion (1214); Fracture (1260); Material Fragmentation (1261)
Patient Problem Reaction (2414)
Event Type  Injury  
Manufacturer Narrative
This report is based on allegations set forth in plaintiff¿s complaint, and the allegations contained therein are unverified.The product identification necessary to review manufacturing history was not provided.Current information is insufficient to permit a conclusion as to the cause of the event.Device product code - ni.Expiration date - ni.Device code - 2616 malposition of device.Manufacture date ¿ ni.
 
Event Description
Legal counsel for patient reported that the patient's left hip was revised two years post-implantation due to fragmentation and failure of polyethylene, fractured locking ring, evidence of a dome screw wear on the backside of the liner, metallosis stained synovial tissue in the periacetabular region and malposition of the dome screw.
 
Manufacturer Narrative
Upon reassessment of the reported event, it was determined an mdr should not have been filed under the current mfr number.This event will be reported under mfr number 1822565.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN HIP
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER, INC.
1800 west center street
warsaw IN 46580
Manufacturer (Section G)
ZIMMER, INC.
1800 west center street
warsaw IN 46580
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key6309472
MDR Text Key66763072
Report Number0001822565-2017-00531
Device Sequence Number1
Product Code JDI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PNI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation Attorney
Type of Report Initial,Followup
Report Date 03/08/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/09/2017
Initial Date FDA Received02/07/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received03/10/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age60 YR
-
-