• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. ATTUNE FB PS ARTIC SURF SZ8; KNEE INSTRUMENT/TRIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. ATTUNE FB PS ARTIC SURF SZ8; KNEE INSTRUMENT/TRIAL Back to Search Results
Catalog Number 254500508
Device Problem Component Missing (2306)
Patient Problem No Information (3190)
Event Date 01/12/2017
Event Type  malfunction  
Manufacturer Narrative
This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Missing balseal.
 
Manufacturer Narrative
Device available for evaluation.Examination of the returned device confirms the reported event of a missing balseal on the smaller post of the trial.Hhe/qrb (quality review board) (b)(4) recommended a device correction which was initiated on (b)(6) 2015.Capa-(b)(4) determined the likely root cause to be related to misuse, and requires mandatory sales training on proper use and application of the device by depuy sales consultants.Capa-(b)(4) will also monitor the mandatory field training.The need for further corrective action was not indicated.Continue to monitor disassociation events per post market surveillance sep-(b)(4).Depuy considers the investigation closed at this time.Should the additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE FB PS ARTIC SURF SZ8
Type of Device
KNEE INSTRUMENT/TRIAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC.
700 orthopaedic drive
warsaw IN 46582
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC.
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
chad gibson
700 orthopaedic drive
warsaw, IN 46581
5743725905
MDR Report Key6310766
MDR Text Key66821519
Report Number1818910-2017-12130
Device Sequence Number1
Product Code MBH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K140881
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Health Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date 01/13/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number254500508
Device Lot NumberMVMBTR070
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/13/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/13/2017
Initial Date FDA Received02/08/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received02/17/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/23/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberZ-2137/2146-2015
Patient Sequence Number1
-
-