MAQUET CARDIOPULMONARY AG RESERVOIR, BLOOD, CARDIOPULMONARY BYPASS; FILTER, BLOOD, CARDIOPULMONARY BYPASS, ARTERIAL LINE Back to Search Results
Model Number VKMO10000
Device Problem Occlusion Within Device (1423)
Patient Problems Hyperglycemia (1905); Thrombus (2101)
Event Date 01/10/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A follow-up medwatch will be submitted when additional information becomes available.
 
Event Description
According to the customer: "this device was used on a neonatal patient, for a tetrology of fallot procedure.Pre-procedurally the patient had act of 600, during bypass 580 anf then at end of procedure 400+.Platelets were normal, no heparin given.Tubing set was blood-primed with no products added.During the case the volume dropped from 250ml by 50ml (which is normal) at which point the perfusion noticed a clot forming within the screen of the reservoir.The clot was of pasty consistency.There was no clot found on the arterial outlet or within the arterial filter.The hospital are not sure what caused the clot, but felt it necessary to report, nonetheless.Unfortunately the device was discarded at the end of the procedure.The procedure was completed normally, there was no effect to the patient, who was discharged normally from icu." (b)(4).
 
Manufacturer Narrative
(b)(4).The visual investigation of the received pictures showed that the reservoir shows clotting.The sample was discarded by the hospital.Clotting is a known phenomenon to maquet and has been thoroughly investigated in a previous complaints,4 additional complaints were recorded.Device history record of the lot has been investigated and no abnormality was found.An internal clinical assessment was performed based on the received complaint report.Though the device of complaint was coated clotting may occur when the perfusion takes a steady course over a longer period of time.When the preload and afterload conditions solely changes slightly and the suction/vent volume also maintaining steady.The initial act level was high but the report mentions that no additional heparin given.It is unknown what exactly 400+act mean and as a combination of a steady perfusion with a slightly low anticoagulation a promotion of coagulation could have occurred.The high activated suction blood probably increased the progression of the clotting.Furthermore the report states the volume dropped during the case and so the viscosity of the perfusate may have increased and also contributed to the clotting.It is not reported and remains unclear if not coagulatory substances,such as blood substitution products have been applied during the case,which likely could have promoted the finding.The photographs unveil significant filter thrombosis,but a conclusive assessment is barely possible.Even with more information from the hospital it is hard to determine.Taking the accessible information into consideration it is deemed to be not necessarily hence a sole device mal-performance.The exact root cause of the failure could not be determined.This data will be handled through a designated maquet trending process.
 
Event Description
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESERVOIR, BLOOD, CARDIOPULMONARY BYPASS
Type of Device
FILTER, BLOOD, CARDIOPULMONARY BYPASS, ARTERIAL LINE
Manufacturer (Section D)
MAQUET CARDIOPULMONARY AG
rastatt
GM 
Manufacturer (Section G)
BERND RAKOW
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt
GM  
Manufacturer Contact
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt 
4972229321
MDR Report Key6311659
MDR Text Key67228870
Report Number8010762-2017-00037
Device Sequence Number1
Product Code DTM
Combination Product (y/n)N
PMA/PMN Number
K141432
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 07/11/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2018
Device Model NumberVKMO10000
Device Catalogue Number70105019
Device Lot Number92187533
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/11/2017
Initial Date FDA Received02/08/2017
Supplement Dates Manufacturer Received01/11/2017
Supplement Dates FDA Received07/11/2017
Date Device Manufactured02/01/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-