Model Number SC-2317-50 |
Device Problems
Fracture (1260); High impedance (1291); Kinked (1339)
|
Patient Problems
Device Overstimulation of Tissue (1991); Inadequate Pain Relief (2388)
|
Event Date 01/19/2017 |
Event Type
Injury
|
Manufacturer Narrative
|
Additional suspect medical device components involved in the event: model#: sc-1132, serial#: (b)(4), description: precision spectra implantable pulse generator.Model#: sc-2317-50, serial#: (b)(4), description: infiniontm cx 50 cm lead.Model#: sc-4316, lot#: 18390607, description: next generation anchor kit-sterile.
|
|
Event Description
|
A report was received that the patient was experiencing inadequate stimulation.High impedances were noted.X-ray result confirmed that the lead was kinked.The physician suspected that the lead was fractured which caused the patient being shocked when turning on the device.The patient underwent a revision procedure wherein the ipg, leads, and clik anchor were replaced.
|
|
Manufacturer Narrative
|
Additional information was received that the patient was doing well postoperatively.The explanted devices were not returned to bsn as they were discarded by the medical facility.It is indicated that the devices will not be returned for evaluation; therefore a failure analysis of the complaint devices could not be completed.A review of the device history records will be conducted.If there is any further relevant information from that review, a supplemental med watch will be filed.
|
|
Event Description
|
A report was received that the patient was experiencing inadequate stimulation.High impedances were noted.X-ray result confirmed that the lead was kinked.The physician suspected that the lead was fractured which caused the patient being shocked when turning on the device.The patient underwent a revision procedure wherein the ipg, leads, and clik anchor were replaced.
|
|
Search Alerts/Recalls
|