Model Number RVDR01 |
Device Problems
Inappropriate or Unexpected Reset (2959); Missing Test Results (3267)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 11/01/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product event summary: the device was not returned for analysis.However, performance data collected from the device was received and analyzed.Analysis of the device memory indicated the battery measurement was not available.Battery voltage not available due to bit flip in lead impedance/battery voltage trend data.
|
|
Event Description
|
It was reported that upon interrogation of the device, the battery voltage reading was not available and was not available upon re-interrogation.It was noted that the patient has been undergoing radiation therapy.A review of the product performance data was able to find the battery voltage.The device remains in use.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was further reported that again, the device battery voltage was not available upon interrogation several months after the first instance.It was noted that the prior instance was due to a partial power on reset (por).It was noted that an ¿initialize data¿ was not performed six month ago.The initialize data was performed and the device remains in use.
|
|