Model Number CENTRALINK DATA MANAGEMENT SYSTEM
Device Problem Device Issue (2379)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/16/2017
Event Type  malfunction  
Manufacturer Narrative
The customer contacted a siemens customer care center (ccc).The ccc contacted the regional service center (rsc).The rsc found the mispl filter to be incorrect.The rsc removed the mispl filter.The customer ran a patient sample and confirmed this corrected the issue.The cause of the cardiac troponin sample not being flagged as critical was due to the mispl filter not properly being set.The instrument is performing according to specifications.No further evaluation of the device is required.
 
Event Description
The customer reported that critical cardiac troponin test result was not properly flagged on their centralink data management system.The physician(s) as a result, were not called to review the result.The initial result was reported out to the customer.The customer did not repeat the sample.There are no known reports of patient intervention or adverse health consequences due to the result not being flagged as critical.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRALINK DATA MANAGEMENT SYSTEM
Type of Device
DATA MANAGER
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
511 benedict ave.
tarrytown NY 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICSMANUFACTURING LTD
registration #: 8020888
chapel lane
swords, co, dublin 10040
EI   10040
Manufacturer Contact
timothy rice
511 benedict ave
tarrytown, NY 10591
9145242406
MDR Report Key6324359
MDR Text Key67326536
Report Number2432235-2017-00113
Device Sequence Number1
Product Code JQP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Health Professional
Type of Report Initial
Report Date 02/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCENTRALINK DATA MANAGEMENT SYSTEM
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/16/2017
Initial Date FDA Received02/10/2017
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age54 YR
-
-