Catalog Number 9734680
Device Problems Bent (1059); Material Deformation (2976)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/20/2017
Event Type  malfunction  
Manufacturer Narrative
Patient information was not made available from the site.Return requested.Replacement thoracic probe shipped to site 01/20/2017.Medtronic investigation of returned suspect thoracic probe finds that the tip of the thoracic probe was twisted.Reported issue confirmed.Bent instrument tip.Failure: physical damage.Hardware investigation was completed.This issue was found related to a hardware issue and was documented in a medtronic hardware anomaly tracking database.
 
Event Description
A medtronic representative received a report from a site that while in a t10-l2 fusion spine procedure, the thoracic probe was being used, the tip became visibly bent.No further details regarding the damage, or how it occurred, were provided.The surgeon opted to continue and completed the procedure with the use of the navigation system.There was no delay of therapy reported.There was no impact on patient outcome.
 
Manufacturer Narrative
Patient demographics provided.
 
Manufacturer Narrative
There was a reported delay to the procedure of less than 1 hour due to this issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAV THORACIC PROBE TIP
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
7208902187
MDR Report Key6329142
MDR Text Key67441476
Report Number1723170-2017-00742
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00613994870063
UDI-Public00613994870063
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K124004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup,Followup
Report Date 07/05/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Catalogue Number9734680
Device Lot Number160920
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/31/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/20/2017
Initial Date FDA Received02/14/2017
Supplement Dates Manufacturer ReceivedNot provided
06/07/2017
Supplement Dates FDA Received03/28/2017
07/05/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/20/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age62 YR
Patient Weight76
-
-