An event regarding ¿elevated metal ions and cobaltism¿ involving a metal head was reported.The event was not confirmed.Visual, dimensional, functional inspection, and material analysis were not performed as the item was not returned.A review of the provided medical records and x-rays by a clinical consultant indicated: ¿based upon the information available for review, there is no confirmation of the analysis of the surgeon implicating the head/neck corrosion as the cause of this clinical situation.¿ review of the device history records indicates devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for this lot.The event could not be confirmed nor the root cause determined based on the clinician¿s report that ¿based upon the available clinical information for review, there is no confirmation of the analysis of the surgeon implicating the head/neck corrosion as the cause of this clinical situation.¿ the device was not returned or identified in this complaint.If the devices and/or additional information are received, this investigation will be reopened and re-evaluated.
|