Product analysis was completed on the returned generator.Visual examination showed only observations consistent with the explant process, including evidence of use of electrocautery through burn marks.No surface abnormalities were noted on this device.With the generator opened, no visual anomalies were found on the printed circuit board assembly.During the bench interrogation, the pulse-disabled and end of service (eos) warnings were set.The pulse-disabled status would not reset.The printed circuit board assembly was subjected to a postburn electrical test.All tests required for proper device function passed.The battery measured 1.803 volts, confirming an eos condition.Although it is possible the generator was previously at an neos condition before being hit with electrocautery, the ultimate cause of the eos condition was determined to be normal current consumption.The internal device data showed that 114.370% of the expected battery capacity had been consumed.Review of the internal device revealed no anomalies.Other than the expected eos condition, there were no performance or any other type of adverse conditions found with the pulse generator.
|