• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES DR CUSTOM DEFINED DISPOSABLE PRESSURE TRANSDUCER; TRANSDUCER, PRESSURE, CATHETER TIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES DR CUSTOM DEFINED DISPOSABLE PRESSURE TRANSDUCER; TRANSDUCER, PRESSURE, CATHETER TIP Back to Search Results
Model Number VO3304TTPL01
Device Problem Particulates (1451)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/02/2017
Event Type  malfunction  
Manufacturer Narrative
The device evaluation is anticipated.However the complaint could not be confirmed without the completion of the product evaluation.A supplemental report will be forthcoming with the evaluation and device history results when received.
 
Event Description
It was reported that an unknown dust like particulate was found inside the pressure tubing during priming before use.It was indicated that the particulate may have moved due to the saline inside of the tubing.Occurrence date is unknown.There were no patient complications reported.Inquired of patient demographics.Unable to be obtained.
 
Manufacturer Narrative
We received one triple flothru dpt-vamp flex kit with an iv set and pressure tubing for examination.The reported event of ¿unknown dust like particulate was found inside of the pressure tubing¿ was not confirmed.The only mark/stain noticed on the tubing is where the customer marked in black ink the location of the particulate.However, no visible particulates was observed from the location during a visual examination.The kit was flushed continuously for 5 minutes, but no visible particulates were flushed out from the kit.The reported event was not confirmed.Although the cause of the complaint could not be determined, there was no indication of a manufacturing defect noted during the analysis.With dpt sets, it is common for the clinician to check for particulates while priming the line for use.Additionally, these products are used by highly trained clinicians, experienced in identifying and mitigating any hazards that arise.Invasive procedures involve some patient risks.Although serious complications are relatively uncommon, the physician is advised to consider the potential benefits in relation to the possible complications.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.No corrective or preventative actions are required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUSTOM DEFINED DISPOSABLE PRESSURE TRANSDUCER
Type of Device
TRANSDUCER, PRESSURE, CATHETER TIP
Manufacturer (Section D)
EDWARDS LIFESCIENCES DR
parque industrial de itabo
carr. sanchez km 18.5
haina, san cristobal
DR 
Manufacturer (Section G)
EDWARDS LIFESCIENCES DR
parque industrial de itabo
carr. sanchez km 18.5
haina, san cristobal
DR  
Manufacturer Contact
brian hurd-servin
1 edwards way
irvine, CA 92614
9492506423
MDR Report Key6355205
MDR Text Key68216275
Report Number2015691-2017-00476
Device Sequence Number1
Product Code DXO
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/02/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/01/2018
Device Model NumberVO3304TTPL01
Device Lot NumberHL0891MT
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/13/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/02/2017
Initial Date FDA Received02/23/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/03/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-