(b)(4).The analysis results found that the el5ml device was returned with no damage in the external components.In an attempt to replicate the reported incident, the instrument was tested for functionality.During the analysis, the device was cycled and it fed and formed 9 conforming clips.Upon testing, the jaws open and close without any difficulties.In addition, the device locked out as intended.No conclusion could be reached as to what may have caused the reported incident.The batch history record was reviewed and identified two class ii defects during in process related with anti-backup failures.When this occurs, our quality system documents the necessary actions to ensure final product quality.Batch was inspected and released without failures.The final quality release criteria were met before this batch was released for distribution.In addition, no protocols or ncr related to the complaint, were found during the manufacturing process.
|