• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC UNIVERSA FIRM URETERAL STENT SET; FAD STENT, URETERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC UNIVERSA FIRM URETERAL STENT SET; FAD STENT, URETERAL Back to Search Results
Model Number N/A
Device Problems Break (1069); Material Separation (1562); Stretched (1601)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/29/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The event is currently under investigation.
 
Event Description
The reporter stated they received a generalized complaint from the physician who stated she was experiencing some issues while using the 6 french universa firm ureteral stent set with hydrophilic coating in different lengths.The physician stated that the tether on the stents was stretching and becoming too thin, it was curling and at times it would break or disconnect from the stents.It is not clear how many actual devices this happened to.There were no unintended sections of the device that remained inside of the patient¿s body, nor did the patient experience any adverse effects due to this occurrence.The customer indicated there would be no photos sent of the device, nor return of the device.No further information was provided.
 
Manufacturer Narrative
Investigation - evaluation.A review of documentation, manufacturing instructions, instructions for use (ifu), and quality control data was conducted during the investigation.Based on the investigation evaluation, there is no indication that a design or process related failure mode contributed to this event.Current controls for manufacturing are in place to assure functionality; device integrity prior to shipping.Review of production and quality documentation did not observe any specific issues with current manufacturing or quality controls that may have contributed to this incident.A lot number was not provided, therefore a review of non-conformances and complaint history search could not be completed.As no sample was returned and no lot numbers were provided a root cause cannot be determined for the reported event.We will notify the appropriate personnel and continue to monitor for similar complaints.As per the risk assessment no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERSA FIRM URETERAL STENT SET
Type of Device
FAD STENT, URETERAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
larry pool
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key6357151
MDR Text Key68363229
Report Number1820334-2017-00317
Device Sequence Number1
Product Code FAD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K961446
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 11/03/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue NumberUFH-626
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/30/2017
Initial Date FDA Received02/24/2017
Supplement Dates Manufacturer Received10/18/2017
Supplement Dates FDA Received11/03/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-