• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH UNKNOWN CURVED CUP IMPACTOR; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH UNKNOWN CURVED CUP IMPACTOR; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 1440-2010
Device Problems Break (1069); Material Integrity Problem (2978)
Patient Problems No Consequences Or Impact To Patient (2199); No Information (3190)
Event Date 01/26/2017
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of the investigation.
 
Event Description
During a primary right hip surgery the surgeon held up the curved cup impactor and it snapped in half.
 
Manufacturer Narrative
An event regarding a fractured da instrument impaction handle was reported.The event was confirmed.Method & results: device evaluation and results: visual inspection was performed on the returned device by the materials analysis engineer, this concluded "after examining the device, it was determined the curved cup impactor fractured due to an overload condition".Medical records received and evaluation: not performed as no adverse consequences to the patient were noted.Device history review: review indicated all devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the reported lot.The investigation concluded that the device fractured due to an overload condition.
 
Event Description
During a primary right hip surgery the surgeon held up the curved cup impactor and it snapped in half.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN CURVED CUP IMPACTOR
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
Manufacturer Contact
rita intorrella
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key6357520
MDR Text Key68595112
Report Number0002249697-2017-00664
Device Sequence Number1
Product Code LPH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K153345
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/30/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number1440-2010
Device Lot NumberDC250087
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/10/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/26/2017
Initial Date FDA Received02/24/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received03/30/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/15/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age84 YR
-
-