Model Number M00545000 |
Device Problem
Break (1069)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 01/17/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Although the suspect device has been received, the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
|
|
Event Description
|
It was reported to boston scientific corporation that an rx cytology brush was used during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2017.According to the complainant, the brush ¿went¿ broke during the procedure.There were no known patient complications reported as a result of this event.
|
|
Manufacturer Narrative
|
Investigation results: visual evaluation of the returned device revealed that the catheter was kinked in multiple locations throughout.The pull wire was cut at the distal end and the bristled portion of the brush was not present and was not returned for evaluation.There were residues found on the device which indicate use and handling.Functional analysis showed that the handle was retracted, however, the pull wire failed to retract.The device was disassembled and found out that the pull wire was broken at the hypotube joint.Due to anatomical/procedural factors encountered during the procedure; performance of the device was limited.Therefore, the most probable root cause classification for the reported failure is operational context.A review of the device history record (dhr) was performed and no deviations were found.
|
|
Event Description
|
It was reported to boston scientific corporation that an rx cytology brush was used during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2017.According to the complainant, the brush ¿went¿ broke during the procedure.There were no known patient complications reported as a result of this event.
|
|