Catalog Number HP054
Device Problem Loss of Power (1475)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/19/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device was received with the nose cone cracked and the coating material of the housing flaking off.The loose particles on the surface are aluminum oxide from the base material.This issue does not affect handpiece performance.The handpiece was connected to a generator, evaluated with a test instrument and was found to be functional.The handpiece was fully functional and performed as per requirements.The handpiece was disassembled to inspect the internal components, no evidence was found that could have caused the reported activation issues.No conclusion could be reached as to what caused this issue.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
 
Event Description
It was reported that during a laparoscopic low anterior resection, the generator was powered off suddenly three times in a row.The hand piece was replaced to complete the case.There were no adverse consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo PR 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC
475 calle c
guaynabo PR 00969
Manufacturer Contact
milton garrett
4545 creek road ml 120a
cincinnati, OH 45242
5133378865
MDR Report Key6363887
MDR Text Key68836387
Report Number3005075853-2017-01103
Device Sequence Number1
Product Code LFL
Combination Product (y/n)N
PMA/PMN Number
K002906
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 01/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberHP054
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/15/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/15/2017
Initial Date FDA Received02/28/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/08/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
GENERATOR
-
-