COVIDIEN KANGAROO EPUMP - NEW; ENTERAL FEEDING PUMP Back to Search Results
Model Number 382400
Device Problem Failure to Power Up (1476)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
An investigation is currently underway.Upon receipt, the results will be forwarded.
 
Event Description
It was reported to covidien on (b)(6) 2016 that a customer experienced an issue with an enteral feeding unit.The customer states that the unit has no power.Upon triage, the unit was found to have intermittent power.
 
Manufacturer Narrative
An evaluation of the kangaroo pump was performed for the reported condition that the unit has no power.The unit was triaged and the reported issue was confirmed.Liquid ingress caused the pcba to corrode, and is the result of customer misuse.All device history records are reviewed for quality inspections and parameter compliance prior to releasing the product for shipment.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KANGAROO EPUMP - NEW
Type of Device
ENTERAL FEEDING PUMP
Manufacturer (Section D)
COVIDIEN
15 hampshire street
mansfield
Manufacturer (Section G)
COVIDIEN
15 hampshire street
mansfield MA
Manufacturer Contact
edward almeida
15 hampshire street
mansfield, MA 02048
5084524151
MDR Report Key6364340
MDR Text Key68979221
Report Number1282497-2017-00240
Device Sequence Number1
Product Code LZH
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number382400
Device Catalogue Number382400
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/01/2016
Initial Date FDA Received02/28/2017
Supplement Dates Manufacturer Received06/01/2016
Supplement Dates FDA Received10/12/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-