• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THORATEC SWITZERLAND GMBH THORATEC CENTRIMAG PRIMARY CONSOLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THORATEC SWITZERLAND GMBH THORATEC CENTRIMAG PRIMARY CONSOLE Back to Search Results
Catalog Number 201-90411
Device Problem Overheating of Device (1437)
Patient Problems Complaint, Ill-Defined (2331); No Known Impact Or Consequence To Patient (2692)
Event Date 02/02/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The primary console is not a single use device.The approximate age of the device from the date of manufacturer is 1 year.Additional information has been requested, but has not been provided.A supplemental report will be submitted when the manufacturer's investigation is completed.
 
Event Description
The patient was implanted with an extracorporeal circulatory support device.It was reported that the motor "overheated" while connected to the primary console.The device was manually turned off before any alarms could be recorded, and the primary console was replaced.The primary console was supporting a patient at the time of the event, and no patient harm was reported.Additional information has been requested, but has not been provided.
 
Manufacturer Narrative
The primary console was not returned for analysis.As a result, the reported event could not be confirmed and a root cause could not be conclusively determined.A review of the device history records revealed no deviations from manufacturing or quality assurance specifications.No further information was provided.The manufacturer is closing the file on this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THORATEC CENTRIMAG PRIMARY CONSOLE
Type of Device
CENTRIMAG PRIMARY CONSOLE
Manufacturer (Section D)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich 8005
SZ  8005
Manufacturer (Section G)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer Contact
bob fryc
6101 stoneridge dr.
pleasanton, CA 94588
7818528390
MDR Report Key6369142
MDR Text Key68762880
Report Number2916596-2017-00416
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 04/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2018
Device Catalogue Number201-90411
Other Device ID NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/02/2017
Initial Date FDA Received03/01/2017
Supplement Dates Manufacturer Received03/30/2018
Supplement Dates FDA Received04/24/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/01/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-