• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ADVANCED STERILIZATION PRODUCTS STERRAD® CYCLESURE® 24 BIOLOGICAL INDICATOR; INDICATOR, BIOLOGICAL (FRC)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADVANCED STERILIZATION PRODUCTS STERRAD® CYCLESURE® 24 BIOLOGICAL INDICATOR; INDICATOR, BIOLOGICAL (FRC) Back to Search Results
Catalog Number 14324
Device Problems Device Operates Differently Than Expected (2913); Environmental Particulates (2930)
Patient Problem No Patient Involvement (2645)
Event Date 02/10/2017
Event Type  malfunction  
Manufacturer Narrative
The correct brand name is sterrad® sealsure chemical indicator tape.The correct common device is indicator, chemical.The correct catalog number is 14202.
 
Event Description
A customer reported the sterrad® sealsure chemical indicator tape did not change color correctly after a completed sterrad® nx cycle.The affected load was reprocessed.There was no report of infection, injury or harm to patient(s) associated with this issue.Although there is no report of patient injury or harm and no prior incidents have resulted in serious injury, advanced sterilization products (asp) has determined in this situation sterility cannot be assured.Therefore, as a matter of policy asp had decided to report all incidents of sterrad® sealsure chemical indicator tape not changing color correctly.(b)(4).This is two of two 3500a reports being submitted for this product malfunction.Please reference manufacturer report numbers: 2084725-2017-00108 and 2084725-2017-00109.
 
Manufacturer Narrative
Asp investigation summary: the investigation included a review of the device history record (dhr), trending of lot number, visual analysis, retains testing and system risk analysis (sra).The dhr, trending analysis by lot number, visual analysis and retains testing was not reviewed since there is clear evidence that user error was determined to be the cause of the issue.The sra indicates the risk associated with a quality problem with no impact on safety is "low." the concomitant sterrad® nx was tested by a field service engineer.The unit was in working condition and no maintenance was required.The assignable cause of the issue can be attributed to user error as the customer indicated they had placed towels in the affected load.Per the instructions for use (ifu), it states that any materials that would absorb liquids, such as cellulose, cotton, paper or cardboard, linens, and towels should not be processed in the unit.The customer was advised of proper load preparation.The issue will continue to be tracked and trended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERRAD® CYCLESURE® 24 BIOLOGICAL INDICATOR
Type of Device
INDICATOR, BIOLOGICAL (FRC)
Manufacturer (Section D)
ADVANCED STERILIZATION PRODUCTS
33 technology drive
irvine CA 92618
Manufacturer (Section G)
ADVANCED STERILIZATION PRODUCTS
33 technology drive
irvine CA 92618
Manufacturer Contact
joaquin kurz
irvine, CA 92618
9497893837
MDR Report Key6375021
MDR Text Key68978272
Report Number2084725-2017-00109
Device Sequence Number1
Product Code FRC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K994055
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number14324
Other Device ID Number14324
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/10/2017
Initial Date FDA Received03/02/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/10/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-