DATEX-OHMEDA, INC. AESTIVA 7900; ANESTHESIA GAS MACHINE Back to Search Results
Model Number 1006-9305-000
Device Problem Computer Software Problem (1112)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/01/2017
Event Type  malfunction  
Manufacturer Narrative
A gehc service representative performed a checkout of the equipment and was not able to confirm the reported complaint, however the error was discovered in the error logs.The central processing unit was replaced and the unit was returned to service.Patient information could not be obtained after multiple attempts.Attempts were made as follows: 2/24/17 phone call, 2/28/17 phone call, 3/1/17 phone call.Mfg date: date of device manufacture year is 2004.The month of manufacture was unavailable at time of mdr filing.
 
Event Description
The hospital reported the unit had watchdog software failure during use, this failure would have resulted in an inability to mechanically ventilate.There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AESTIVA 7900
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda drive
madison WI
Manufacturer Contact
john szalinski
540 w. northwest highway
barrington, IL 
MDR Report Key6376529
MDR Text Key69018341
Report Number2112667-2017-00433
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K023366
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 03/03/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1006-9305-000
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/03/2017
Initial Date FDA Received03/03/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/01/1970
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-