• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 640G INSULIN PUMP; INSULIN INFUSION PUMP / SENSOR AUGMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 640G INSULIN PUMP; INSULIN INFUSION PUMP / SENSOR AUGMENTED Back to Search Results
Model Number MMT-1712P
Device Problems Crack (1135); Device Dislodged or Dislocated (2923)
Patient Problem Hyperglycemia (1905)
Event Date 01/25/2016
Event Type  malfunction  
Manufacturer Narrative
The insulin pump was unable to perform displacement test due to missing retainer.The pump was received with minor scratched lcd window, scratched case, pillowing keypad overlay and missing reservoir tube o-ring.No crack was found around reservoir compartment.The insulin pump involved in this event is the 640g insulin infusion pump, which is not marketed in the united states.However, the device is similar to the paradigm real-time insulin infusion pump, which is marketed in the united states.
 
Event Description
It was reported of damage on the insulin pump.The customer's blood glucose level was 246 mg/dl at the time of the incident.The customer stated that there was a crack around the reservoir compartment.The customer stated that the reservoir o-ring was dislodged.The customer stated that the pump was not dropped or bumped.The product was returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
640G INSULIN PUMP
Type of Device
INSULIN INFUSION PUMP / SENSOR AUGMENTED
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer Contact
gerwin de graaff
18000 devonshire street
northridge, CA 91325-1219
8185764805
MDR Report Key6377710
MDR Text Key69342049
Report Number3004209178-2017-93787
Device Sequence Number1
Product Code OYC
Combination Product (y/n)N
Reporter Country CodeSP
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 02/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-1712P
Device Catalogue NumberMMT-1712P
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/12/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/10/2017
Initial Date FDA Received03/03/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-