Model Number N/A
Device Problem Fracture (1260)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/04/2017
Event Type  malfunction  
Manufacturer Narrative
Current information is insufficient to permit a conclusion as to the cause of the event.Review of device history records show the lot released with no recorded anomaly or deviation.In the package insert, under care and handling, it states: "surgical instruments are subject to wear with normal usage.Instruments, which have experienced extensive use or excessive force, are susceptible to fracture.Surgical instruments should only be used for their intended purpose." (b)(4).
 
Event Description
It was reported that the clamp fractured during the tightening of the device.Device had been in use for 3-4 months.Another instrument was used to finished the surgery non-optimally.No further information is available at this time.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Customer has indicated that the product will not be returned to zimmer biomet for investigation, as it has been discarded.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Manufacturer Narrative
Reported event was confirmed by review of photographs provided.Device history record (dhr) was reviewed and no discrepancies were found.Review of the complaint history determined that no further action(s) is/are required.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BMP CABLE SLEEVE/GRIP CRIMPER
Manufacturer (Section D)
BIOMET TRAUMA
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key6377799
MDR Text Key69267804
Report Number0001825034-2017-01356
Device Sequence Number1
Product Code HXQ
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 09/25/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberN/A
Device Catalogue Number498003
Device Lot Number765850
Other Device ID NumberSEE H10 NARRATIVE
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/07/2017
Initial Date FDA Received03/03/2017
Supplement Dates Manufacturer ReceivedNot provided
09/22/2017
Supplement Dates FDA Received04/03/2017
09/25/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/26/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Age91 YR
-
-