• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY AG QUART ARTERIAL FILTER WITH SOFTLINE COATING; FILTER, BLOOD, CARDIOPULMONARY BYPASS, ARTERIAL LINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY AG QUART ARTERIAL FILTER WITH SOFTLINE COATING; FILTER, BLOOD, CARDIOPULMONARY BYPASS, ARTERIAL LINE Back to Search Results
Model Number BO-H 20306
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/22/2017
Event Type  malfunction  
Manufacturer Narrative
Maquet getinge group is aware of similar complaints from this product.Similar products, showing a similar malfunction, have been tested.We tested the returned filter in our laboratory, following our standard process, on tightness.Therefore, the quart was tested at below water in a water bath and was then pressurized with compressed air (0.3 bar).Due to the air escaping (visible as air bubbles in water) a leakage in the welding cover has been found.The leakage at the connection between cover and filter body can be confirmed.Most possible root cause could be the bad welding between cover and filter body.Since the reported failure did not contribute to a death or serious injury no corrective action is needed.In addition at this time it cannot be concluded that this is an systemic error.This data will be handled through a designated maquet trending process.If a trend occurs, it will be escalated to quality assurance management for review and determination if further investigation is necessary.
 
Event Description
Customer reported that "during the product evaluation test with priming solution, the customer found that the solution was leaking" no known consequences to the patient.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUART ARTERIAL FILTER WITH SOFTLINE COATING
Type of Device
FILTER, BLOOD, CARDIOPULMONARY BYPASS, ARTERIAL LINE
Manufacturer (Section D)
MAQUET CARDIOPULMONARY AG
rastatt
GM 
Manufacturer (Section G)
BERND RAKOW
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt
GM  
Manufacturer Contact
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt 
4972229321
MDR Report Key6383658
MDR Text Key69573327
Report Number8010762-2017-00074
Device Sequence Number1
Product Code DTM
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K090518
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/01/2017
Device Model NumberBO-H 20306
Device Catalogue Number701048784
Device Lot Number70104216
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/21/2017
Initial Date FDA Received03/07/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-