The customer stated that a patient who performs testing with coaguchek xs meter serial number (b)(4) had an ischemic event in the eye on (b)(6) 2017 and was brought by ambulance to the hospital.On (b)(6) 2017, the patient was tested in parallel on the meter and with the stago sta-r laboratory method.The results from the parallel testing are as follows: (b)(6) 2017: meter result = 3.6 inr, sta-r result = 2.3 inr; (b)(6) 2017: meter result = 3.1 inr, sta-r result = 2.38 inr; (b)(6) 2017: meter result = 2.6 inr, sta-r result = 2.12 inr; (b)(6) 2017: meter result = 2.7 inr, sta-r result = 2.1 inr.As the results from the meter were significantly higher, clinicians concluded that the patient's anticoagulant therapy was insufficient and this led to the ischemic event.The physician prescribed a new coaguchek xs meter to the patient.The current condition of the patient was asked for, but is not known.It was also asked if the patient experienced any further adverse events as a result of the ischemia, but this is not known.The patient's therapeutic range is 2.5 - 3.5 inr.The patient is a heart patient and has a mechanical heart valve since 2001.The patient takes sintrom medication daily.The patient's product was requested for investigation.Relevant retention test strips (lot 119116-10) were tested in comparison with the current master lot coaguchek xs pt test strip (lot 124158-80).For this purpose two human blood samples from marcumar donors and two internal reference meters were used.No error messages occurred.Retention samples were within specifications.
|
The investigation results for donor 2 were initially stated incorrectly.The correct information is as follows: donor 2: master lot with reference meter: 3.8 inr, customer strips with retention strip: 3.9 inr.Medical assessment of the event was performed.Anticoagulant therapy with vitamin k antagonists is known to reduce the risk of stroke.However, there are individual factors (mechanical valve, comorbidities) that can contribute or lead to a thrombosis and ischemic event.No product problem was found.
|