Catalog Number 175816
Device Problem Hole In Material (1293)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event Description
It was reported that there was a pinhole in the balloon of the foley catheter.Subsequently, the catheter was replaced.
 
Manufacturer Narrative
Received 1 used silicone catheter with the drainage bag still attached.Per the visual evaluation no defects were found.Per the functional evaluation, the balloon was inflated with 10 cc of a mix of tap water and blue methylene using a syringe, and immediately the water came out due to a balloon pinhole.The sample was observed under 10 x magnification and no particles were found along the balloon area.The balloon active length measured 0.7870¿ on one side and 0.7995¿ on the other side (specification= 0.6 - 0.9 in.).The balloon active length was found to be within specifications.The reported event was confirmed with an unknown cause.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: "recommended inflation capacities 3cc balloon: use 5ml sterile water, 5cc balloon: use 10ml sterile water, 30cc balloon: use 35ml sterile water.Do not exceed recommended capacities.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Visually inspect the product for any imperfections or surface deterioration prior to use." (b)(4).The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event Description
It was reported that there was a pinhole in the balloon of the foley catheter.Subsequently, the catheter was replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX LUBRI-SIL ALL-SILICONE FOLEY CATHETER WITH HYDROGEL COATING, 16 FR. 5ML
Type of Device
FOLEY CATHETER
Manufacturer (Section D)
PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590
km. 7 carretera internacional
nogales, sonora 85621
MX  85621
Manufacturer (Section G)
PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590
km. 7 carretera internacional
nogales, sonora 85621
MX   85621
Manufacturer Contact
amy gravley
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key6385983
MDR Text Key69624207
Report Number1018233-2017-00918
Device Sequence Number1
Product Code EZL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K040504
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other,use
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/30/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/28/2021
Device Catalogue Number175816
Device Lot NumberNGAV1825
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/03/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/14/2017
Initial Date FDA Received03/07/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received03/30/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/29/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-