The delivery pusher was returned with the introducer and dispenser coil.The implant coil was not returned.The resistance of the system was noted to be 46.2 ohms and the unit passed the 4-way check with the v-grip.The hypotube was noted to be kinked at the laser weld connector.There was evidence of glue on the distal body coil.The monofilament was noted to be completely necked down and had a long tail.The monofilament was noted to be stretched and completely necked down, likely due to issues with retracting the stuck implant coil from the microcatheter.The implant, being small, may have gotten stuck in the hub of the microcatheter, causing resistance during advancement, and further manipulation may have caused the glue at the tie knot locations to spread over the pusher body coil.The stuck implant may have gotten stretched and then detached prematurely while pulling the pusher back from the microcatheter, as evidenced by the stretched monofilament.Based on the information provided and the product evaluation, the reported detachment can be confirmed; however, the root cause cannot be determined because the conditions of use cannot be recreated during bench testing.
|
It was reported that during a coil embolization treatment, the coil did not advance in the microcatheter, then the coil unexpectedly detached in the microcatheter.Pliers were used to remove the coil.The entire coil was removed from the patient.There was no reported patient injury.The current patient's status is reported to be "fine.".
|