The scope was returned to olympus for evaluation.The evaluation confirmed the reported scope damage.A visual inspection noted the bending section broken.It was also noted that a sharp metal pin was found protruding through the bending section cover approximately 4mm from the proximal end of the bending section adhesive.In addition, the bending section cover was removed and found the skeleton broken.Foreign materials were also noted.The scope was serviced and returned to the user facility.Based on the evaluation findings, the cause of the reported scope damage is likely attributed to user handling and improper maintenance of the device.The instruction manual for use states, ¿do not attempt to bend or twist the endoscope¿s insertion section with excessive force.The insertion section may be damaged.Do not twist or bend the bending section with your hands.Equipment damage may result.¿.
|