Model Number MERGE HEMODYNAMICS 9.40.3 PATCH 1
Device Problems Failure to Cycle (1142); Blocked Connection (2888); Data Problem (3196)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/17/2017
Event Type  malfunction  
Manufacturer Narrative
Merge technical support assisted the customer with troubleshooting efforts.It was found that the coil cord and gender changer connector were faulty.The failed hardware was changed out with replacements the customer had on hand; the faulty hardware was not returned to merge healthcare for evaluation.The customer confirmed that the replacement hardware corrected the issue.Device labeling, hemo-5303 user manual - general equipment care, addresses the occurrence for such an issue with statements such as, "inspect overall physical condition of the system components, peripherals, and interconnecting cables.Perform any corrective actions required.".
 
Event Description
Merge hemodynamics monitors, measures, and records physiological data from a human patient undergoing a cardiac catheterization procedure.The system comprises the patient data module and the merge hemodynamics hemo monitor pc.The two units are connected via a serial interface.All vital parameters and evaluations are registered and calculated in the patient data module.This data is then transmitted to the merge hemodynamics hemo monitor pc via the serial interface.All data can be shown and monitored on the merge hemodynamics hemo monitor pc.On (b)(6) 2017, a customer reported to merge healthcare that problems were experienced with the hemo monitor freezing during an emergent procedure.Subsequently, the hemo monitor was rebooted and resulted in a loss of patient monitoring.The delay was ~10-15 minutes while performing troubleshooting efforts; however, the procedure was completed successfully using an external defibrillator equipped with ekg and spo2.With merge hemo not capturing physiological data, there is a potential for incorrect treatment that could cause harm to the patient.However, the customer reported that the procedure was completed successfully using alternate means.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERGE HEMODYNAMICS
Type of Device
PROGRAMMABLE DIAGNOSTIC COMPUTER
Manufacturer (Section D)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer (Section G)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer Contact
michael diedrick
900 walnut ridge drive
hartland, WI 53029
2629123570
MDR Report Key6412273
MDR Text Key70183990
Report Number2183926-2017-00074
Device Sequence Number1
Product Code DQK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K082421
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 02/17/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMERGE HEMODYNAMICS 9.40.3 PATCH 1
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/17/2017
Initial Date FDA Received03/16/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/15/2016