Model Number MERGE HEMODYNAMICS 9.40.1
Device Problems Battery Problem (2885); Power Problem (3010); Noise, Audible (3273)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/20/2017
Event Type  malfunction  
Manufacturer Narrative
Merge technical support shipped replacement hardware, ups tripp lite 1000va (rma #(b)(4)), to the customer on 20feb2017.The faulty unit was returned to merge healthcare on 07mar2017 for evaluation.The results showed that the unit displayed a red led light and the unit was beeping.The batteries were found to be defective.Upon replacing the batteries, the unit passed all testing and was then sent to service stock.It was confirmed that the replacement hardware corrected the customer's problem.
 
Event Description
Merge hemodynamics monitors, measures, and records physiological data from a human patient undergoing a cardiac catheterization procedure.The system comprises the patient data module and the merge hemodynamics hemo monitor pc.The two units are connected via a serial interface.All vital parameters and evaluations are registered and calculated in the patient data module.This data is then transmitted to the merge hemodynamics hemo monitor pc via the serial interface.All data can be shown and monitored on the merge hemodynamics hemo monitor pc.On (b)(6) 2017, a customer reported to merge healthcare that the ups (uninterrupted power supply) displayed a red led light and emitted a beeping noise.The problem was found during lab preparation so there was no patient involvement.If parts of the hemo system become energized, there is a potential for direct harm to the patient and/or user including electrical shock or burns.However, there was no report of harm to anyone at the site as a result of the ups failure.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERGE HEMODYNAMICS
Type of Device
PROGRAMMABLE DIAGNOSTIC COMPUTER
Manufacturer (Section D)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer (Section G)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer Contact
michael diedrick
900 walnut ridge drive
hartland, WI 53029
2629123570
MDR Report Key6416132
MDR Text Key70408848
Report Number2183926-2017-00066
Device Sequence Number1
Product Code DQK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K082421
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 02/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMERGE HEMODYNAMICS 9.40.1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/07/2017
Initial Date Manufacturer Received 02/20/2017
Initial Date FDA Received03/19/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/24/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-