It was reported that during a laparoscopic parastomal hernia procedure, the device stopped working.Cleaned and tightened but would not activate.Tip of the device noted to be loose, but all pieces accounted for.Another like device was used to complete the procedure.There were no adverse consequences for the patient.
|
(b)(4).Evaluation summary was missing from initial 3500a.The remaining blade portion was scratched, and with evidence of contact with metal in or out of the operative field.In addition, the tissue pad detached and not returned but con evidence of body fluids and tissue pad material in the groove section of the clamp arm.During functional testing on gen11 an alert screen was displayed.A probable cause of the device stop activating and display an alert screen is blade damage.The device was disassembled to inspect the internal components and no anomalies were found.Probable causes of blade damage, including breakage, are external contact during pre-op or general use, blade contact with other devices, staples or clips during the procedure or using any means other than the blade wrench to attach or detach the blade.Once minor blade damage has occurred, subsequent activations may increase damage severity and result in alert screens, such as ¿tighten assembly¿, ¿blade error detected¿ or ""relaxed pressure on blade"" followed by a ¿replace instrument¿ screen later in the procedure.The batch history record was reviewed and there were no defects, protocols or ncr(s) found during the manufacturing process related to this complaint.
|