• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HUDSON UP-DRAFT II OPTI-NEB NEBULIZER W/RESERV

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HUDSON UP-DRAFT II OPTI-NEB NEBULIZER W/RESERV Back to Search Results
Catalog Number 1734
Device Problem Inaccurate Delivery (2339)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/02/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device involved in this complaint has not been returned to the manufacturer at the time of this report.The investigation into this complaint is still in progress.
 
Event Description
Customer complaint alleges that the "nebulizer stops nebulizing after a few minutes".Alleged malfunction is reported during use.No report of patient harm or injury.
 
Manufacturer Narrative
(b)(4).The customer returned the actual sample along with 28 unused representative samples.A visual inspection was performed on the samples and no defects or anomalies were found.Approximately 3cc of water was injected into the jar of each kit.The samples were connected to a flowmeter and the airflow was increased to 8 lpm.Functional testing indicated a mist was produced from the chamber of each nebulizer.Each nebulizer began to sputter intermittently after approximately 5 minutes of nebulizing due to low volume of water in the jar to nebulize.The devices were tapped and rotated until they restarted.Based on the functional testing performed, there were no functional issues found with the returned samples.Additional inspection of the jet, jar, and cap were examined under the microscope which revealed no gross imperfections such as flash, debris or concentricity issues.The reported complaint that the nebulizer stops nebulizing after a few minutes could not be confirmed through functional inspection of the returned samples.The devices were tapped and rotated until they restarted.A dhr review was performed with no evidence to suggest a manufacturing related cause.There were no issues found with the returned samples.
 
Event Description
Customer complaint alleges that the "nebulizer stops nebulizing after a few minutes".Alleged malfunction is reported during use.No report of patient harm or injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON UP-DRAFT II OPTI-NEB NEBULIZER W/RESERV
Type of Device
NEBULIZER
Manufacturer (Section D)
TELEFLEX MEDICAL
research triangle park NC
Manufacturer (Section G)
TELEFLEX MEDICAL
parque industrial finsa
nuevo laredo 88275
MX   88275
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
9194334854
MDR Report Key6420273
MDR Text Key70436396
Report Number3004365956-2017-00105
Device Sequence Number1
Product Code CAF
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/02/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Expiration Date11/15/2021
Device Catalogue Number1734
Device Lot Number74L1601536
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/23/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/02/2017
Initial Date FDA Received03/21/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/26/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/09/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-