Catalog Number 000002N
Device Problem Unable to Obtain Readings (1516)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event Description
It was reported that the device was not reading urine volume.Additional information was requested about the patient and the event; however, was unavailable.
 
Manufacturer Narrative
The reported issue of the monitor is not reading the urine was unconfirmed.The criticore was visually examined upon receipt and was found to be in good overall condition.The monitor powered on, and the warranty label is intact.Evaluation of the system revealed there is no root cause as the reported event could not be duplicated.The monitor properly measured all test containers 0 ml, 500 ml and 2100 ml.One empty container was slowly filled and the unit was reading it properly.The coin cell was more than one month old and was replaced.All d cell batteries were replaced.The monitor was serviced and was functioning properly.The criticore was subsequently inspected and tested by bes-dymax in accordance with bes-dymax standard processes and procedures and was found to fully comply with all requirements and specifications.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: "warning: the criticore monitor requires special precautions regarding emc and needs to be installed and put into service according to the emc information provided in the following tables.Portable and mobile rf communications equipment may affect the criticore monitor.Warning: the criticore monitor should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, the criticore monitor should be observed to verify normal operation in the configuration in which it will be used.Warning: do not immerse or submerge the monitor or turn it upside down when cleaning.Caution: use only heavy-duty alkaline d cell batteries.Do not use rechargeable batteries.Do not incinerate batteries.Recycle or dispose of them properly.Contact bard for disposal information.Caution: improper orientation of the batteries within the battery pack can potentially damage the criticore monitor.Caution: state and federal regulations govern the packaging necessary for return of medical product which may have been contaminated.Refer to local, state and federal regulations when packaging the criticore monitor for return.Caution: there are no user serviceable components inside the criticore monitor.The user should not attempt to repair the criticore monitor.To do so may void the warranty and could result in erroneous monitor readings.Caution: use of cables or sensors other than those specified for use with the criticore monitor, except those sold by bard for use as replacement part or repair components, may result in increased emissions or decreased immunity of the criticore monitor.Caution: the criticore monitor should be recycled properly per (b)(4) union directive (b)(4) on waste electronic and electrical equipment, january 27, 2003.Do not dispose with ordinary municipal waste." the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event Description
It was reported that the device was not reading urine volume.Additional information was requested about the patient and the event; however, was unavailable.
 
Manufacturer Narrative
Upon further review, bard medical has determined that this mdr was initially reported in error as this event is not reportable.
 
Event Description
It was reported that the device was not reading urine volume.Additional information was requested about the patient and the event; however, was unavailable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRITICORE FLUID OUTPUT & CORE TEMPERATURE MONITORING SYSTEM - MONITOR
Type of Device
CRITICORE
Manufacturer (Section D)
DYMAX CORP. -2523003
110 marshall drive
warrendale PA 15086
Manufacturer (Section G)
DYMAX CORP. -2523003
110 marshall drive
warrendale PA 15086
Manufacturer Contact
amy gravley
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key6421134
MDR Text Key70699683
Report Number1018233-2017-01224
Device Sequence Number1
Product Code EXS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup
Report Date 12/29/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number000002N
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/03/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/27/2017
Initial Date FDA Received03/21/2017
Supplement Dates Manufacturer ReceivedNot provided
12/29/2017
Supplement Dates FDA Received06/01/2017
12/29/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/08/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-