• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number MT22495
Device Problem No Display/Image (1183)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/28/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The g5 system is associated with product code pqf.
 
Event Description
Dexcom was made aware on (b)(6) 2017 that on (b)(6) 2017, the receiver displays only the back light.No additional patient or event information is available.The complaint device was returned for evaluation.An exterior visual inspection was performed and no defects were found.Functional testing was performed and the test failed.The data log was downloaded and reviewed and a hardware error and screen error alerts were observed.The customer complaint was confirmed.A root cause could not be determined.Based on the investigation results this issue has been upgraded from non-reportable to reportable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of Device
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer (Section G)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer Contact
kipp durbin
6340 sequence drive
san diego, CA 92121
8582000200
MDR Report Key6421809
MDR Text Key70558312
Report Number3004753838-2017-24501
Device Sequence Number1
Product Code MDS
UDI-Device Identifier10386270000153
UDI-Public(01)10386270000153(241)STK-DR-001(10)5211612
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 01/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMT22495
Device Catalogue NumberSTK-DR-001
Device Lot Number5211612
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/21/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/27/2017
Initial Date FDA Received03/21/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/03/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age50 YR
-
-