|
Model Number 52516-15 |
Device Problem
Output Problem (3005)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 03/01/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Lot review: a review of lot# 3333336 showed that (b)(4) units were manufactured, tested, inspected and released in november 2016 citing no exception documents.A review of lot# 3338041 showed that (b)(4) units were manufactured, tested, inspected and released in november 2016 citing no exception documents.
|
|
Event Description
|
Complaint received regarding two 52516-15, triox¿ svo2/cco pa catheter, 8f, 110 cm, j-tip; lot# 3333336 (mfd.11/2016), lot# 3338041 (mfd.11/2016) and lot# 3379096 (mfd.01/2017).Report states:.The laser would not transmit through the catheter; unable to reference light source; and lack of hemodynamic numbers on the display.The latter occurred in different or with different monitors and user.Delay in critical therapy and med/surg intervention required.No adverse patient consequences reported, no baseline changes.
|
|
Manufacturer Narrative
|
Lot review: a review of lot# 3333336 showed that (b)(4) units were manufactured, tested, inspected and released in november 2016 citing no exception documents.A review of lot# 3338041 showed that (b)(4) units were manufactured, tested, inspected and released in november 2016 citing no exception documents.Visual inspection: 3/17/2017 - received two used 52516-15, triox¿ svo2/cco pa catheter, 8f, 110 cm, j-tip; lot# 3379096.The balloons and fiber optics were all function for each catheter.One catheter had a touch contamination shield covering the catheter from the 10 to 80 cm marks.Functional testing: both units were visually inspected no defect was found with the fiber optic.Unit was pre insertion calibrated and no failures were observed.Both units were thruput tested and no defect was found with the catheter fiber optics.The unit functioned per specification.Final analysis summary: the reported complaint of would not transmit through the catheter could not be confirmed.The unit was observed to function per specification.The root cause of the failure could not be determined.Catheter is intended for continuous use up to 72 hours.Icu medical will continue to monitor and trend the reported compliant.
|
|
Event Description
|
Complaint received regarding two 52516-15, triox¿ svo2/cco pa catheter, 8f, 110 cm, j-tip; lot# 3333336 (mfd.11/2016), lot# 3338041 (mfd.11/2016) and lot# 3379096 (mfd.01/2017).Report states:.The laser would not transmit through the catheter; unable to reference light source; and lack of hemodynamic numbers on the display.The latter occurred in different or with different monitors and user.Delay in critical therapy and med/surg intervention required.No adverse patient consequences reported, no baseline changes.
|
|
Search Alerts/Recalls
|
|
|